A Modulus Microplate Fluorometer Method for DNA Quantitation Using Quant-iT™ DNA HS Assay INTRODUCTION Quant-iT™ Assay Kits from Invitrogen are highly sensitive and convenient for DNA, RNA, and protein quantitation in microplate format. The Quant-iT™ DNA High-Sensitivity Assay is highly selective for dsDNA over RNA and tolerates salts, solvents, detergents, protein, and other common contaminants. When used with the Modulus™ Microplate Fluorometer, this assay provides an easy, accurate, and sensitive method for quantitation of DNA in small-volume microplates (200 µL per well). When read in the Modulus™ Microplate Fluorometer in conjunction with the Blue Fluorescence Optical Kit, the signal is linear from 30 pg to 100 ng DNA in 200 µL final volume (Figures 1 and 2).
Figure 2. Samples at low-concentration range analyzed using the Modulus Microplate Fluorometer and the Blue Fluorescence Optical Kit.
1. Reagent Preparation Recommendation Note: Handling, storage and use of the reagent should be performed in accordance with the product information sheet supplied by Invitrogen, Inc. The Quant-iT dsDNA HS Assay is supplied as a 1-mL concentrated dye solution in anhydrous dimethylsulfoxide (DMSO). On the day of the experiment, equilibrate kit contents to room temperature. Prepare a working solution by making a 1:200 dilution of the concentrated Quant-iT dsDNA HS reagent (Component A) in Quant-iT dsDNA HS buffer (Component B). Prepare this solution in a plastic container as the reagent may absorb into glass surfaces. Protect the working solution from light by covering it with foil or placing it in the dark. Note: For best results, use this solution within three hours of preparation. 2. Instrument Set Up
3. Sample Analysis
RESULTS Sensitivity:
typically 150 pg/mL * data determined separately and not shown in this application note. CONCLUSION The Modulus Microplate Reader offers superior sensitivity and dynamic range for detection of common fluorophors used in nucleic acid quantitation such as PicoGreen, RiboGreen, Alexa, and CyDyes. The Modulus Microplate Reader achieves superior performance by using a dedicated fluorescence detector instead of sharing the detector with other detection modes. The individual optical kit with solid-state optics and a powerful wavelength-matched LED delivers excellent sensitivity and dynamic range. The modular approach of the Modulus Microplate Reader allows for instrument capability expansion as needs in the lab change. Luminescence, Fluorescence and/or absorbance detection modules as well as other accessories can be added after the initial purchase. The superior performance, ease of use, and utmost flexibility of the Modulus Microplate Reader make it ideal for today's life science laboratory. ABOUT TURNER BIOSYSTEMS, INC. BioSystems' products may be placed by: Phone: (408)
636-2400 Contact us
via our contact form Mailing Address: TRADEMARKS For research use only. Not for use in diagnostic procedures. Modulus is a trademark of Turner BioSystems, Inc. All other trademarks are the sole property of their respective owners. Quant-iT, PicoGreen, RiboGreen, and Alexa are trademarks of Molecular
Probes, an Invitrogen company. |
|
| Promega Corporation Sunnyvale Office • Detection Instruments 645 N Mary Avenue • Sunnyvale CA 94085 Phone: 408.636.2400 - Toll Free: 888.636.2401 - Fax: 408.737.7919 |
||
|
||