Modulus Microplate Luminometer Method for
Promega Dual-Luciferase® Reporter Assay
INTRODUCTION
The Dual-Luciferase®
Reporter Assay (DLR) from Promega is ideal as a dual genetic reporting
luminescence-based assay system for quantitation of gene expression. Two
different luciferase reporter enzymes, the firefly (Photinus pyralis)
and Renilla (sea pansy or Renilla reniformis), are expressed simultaneously
in each cell creating an efficient means of performing two reporter assays
in a single sample.
The Modulus
Microplate Luminometer with dual injectors provides a convenient, rapid,
and sensitive method for quantifying gene expression when used with Promega's
Dual-Luciferase® Reporter (DLR) Assay kit. The superior sensitivity
and wide dynamic range of the Modulus Microplate Luminometer make
it highly suited for the Dual-Luciferase® Reporter (DLR) Assay application.
The Modulus Microplate Luminometer has a pre-programmed protocol
for quick setup of the Dual-Luciferase® Reporter Assay. Results are
reported in RLUs for each of the luciferase readings and in Ratio (firefly
RLUs/Renilla RLUs) for quick analyses.
The Modulus Microplate Luminometer detects less than 1x10-19 moles
of firefly luciferase enzyme using Luciferase Assay Reagent II (LAR II)
and 1x10-18 moles of Renilla enzyme using Stop & GloÒ Reagent.
Measurements are linear for more than eight and six orders of magnitude
for firefly and Renilla substrates, respectively. All tests were conducted
using purified recombinant firefly luciferase enzyme (Promega Catalog
E1701) and purified Renilla recombinant enzyme (Chemicon Catalog 4400).

Figure 1. Firefly luciferase, 1x10-11 to 1x10-20
moles.

Figure 2. Renilla luciferase,1x10-14 moles.

Figure 3. Ratio of firefly and Renilla luciferase
signals. DLR assay was performed on the
Modulus™ Microplate Luminometer using Promega's Dual-Luciferase® Reporter Assay system with recombinant firefly luciferase and Renilla
luciferase.
MATERIALS REQUIRED
- Modulus
Microplate Multimode Reader
- Modulus Microplate Luminescence Module
- Promega
Dual-Luciferase Reporter® Assay Kit (Promega, Catalog E1980)
- 96-well, white microplates (E&K Scientific, EK-25075)
- P10,
P200 pipette and pipette tips
EXPERIMENTAL
PROTOCOL
1. Reagent
Preparation Recommendation
- Stop &
Glo® Buffer Substrate Solvent: Use as supplied. Store below 25°C.
- Stop &
Glo® Reagent: To make the Stop & Glo® Reagent, dilute the
50x Stop & Glo® Substrate to 1x concentration using Stop &
Glo® Buffer in a glass or siliconized polypropylene tube. Mix by
inversion. Use reconstituted Stop & Glo® Substrate on the same
day it is prepared or store at -20°C for up to two weeks.
- Passive
Lysis Buffer: To make 1x Passive Lysis Buffer, dilute the 5x Passive
Lysis Buffer in DI water with 0.1% Gelatin. Store below 25°C.
Note: Since luciferase activity is temperature dependent, the
temperature of the Luciferase Assay Buffer II and Stop & Glo®
Buffer should be held constant at room temperature while quantifying
luminescence. Reagent stored frozen after reconstitution must be thawed
below 25°C to ensure reagent performance. Mix well after thawing.
The simplest method for thawing is placing the reagent in a water
bath at room temperature.
2. Instrument
Set Up
- From the
Home screen, touch Select Protocol and follow the protocol wizard to
select the preset Dual-Luciferase 2 injectors protocol. Enter the following:
Luminescence; at the Preset tab, select Dual-Luciferase 2 Injectors;
Finish.
- The Instrument
Control screen shows all the reading parameters: integration, each injector
and injection volume, and delay time. All the wells in the plate are
selected to be read. If desired, change the delay and integration time
settings and save the changed protocol under a different protocol name
in the User protocol folder.
- Select
the wells on the Plate Map according to the samples loaded into the
plate.
- Refer
to the on-screen Help topics, Quick Start Guide, or Operating Manual
for detailed instructions.
3. Sample
Analysis
- Touch
the Start icon on the touch screen to begin reading.
- RLU values
measured by the Modulus Microplate Luminometer will appear on
the Results screen of the touch screen display immediately after each
well is measured. Both readings for each well are displayed side-by-side.
Toggle the Ratio icon to alternately view the RLU values and the ratio
of the two (Injector 1 value/Injector 2 value).
- Once the
measurements are complete, data can be transferred to an external computer
for further data analysis in Excel by using the provided USB flash drive.
- Remove
the plate after measurement completion.
- Use the
Reverse Purge function in the Set-Up wizard to return unused reagent
to the bottle.
- Rinse
the injectors using the Flush function in the Set-Up wizard thoroughly
after use.
- Refer
to the Operating Manual for detailed instructions of the use of the
injectors.
RESULTS
Sensitivity:
· < 1x10-19 moles of firefly luciferase using Luciferase Assay
Reagent II (LAR II)
· < 1x10-18 moles of Renilla luciferase using Stop & Glo®
Reagent
Dynamic Range:
· > eight orders of magnitude for firefly luciferase
· > six orders of magnitude for Renilla luciferase
CONCLUSION
The Modulus
Microplate Luminometer offers superior sensitivity and dynamic range for
luminescence detection, such as the luminescence based dual-reporter assays.
The Modulus Microplate Luminometer achieves superior performance
with a combination of unique detection and optical designs, premium components
such as the photomultiplier tube (PMT), low-noise circuitry, and proprietary
dual-masking system.
The modular
approach of the Modulus Microplate Luminometer allows for instrument
capability expansion as needs in the lab change. Fluorescence and/or absorbance
detection modules as well as other accessories can be added after the
initial purchase.
Superior
performance, ease of use, and utmost flexibility of the Modulus
Microplate make it an ideal microplate reader for today's life science
laboratory.
The Modulus Microplate
has passed the Promega Corporation DLReady validation criteria.
CAUTION: The lyophilized Luciferase Assay Substrate contains dithiothreitol (DTT)
and is therefore classified as hazardous. The reconstituted reagent is
not known to present any hazards as the concentration of DTT is less than
1%. However, we recommend the use of gloves, lab coats, and eye protection
when working with these or any chemical reagents. Promega and Turner BioSystems
assume no liability for damage resulting from handling or contact with
these products.
TURNER
BIOSYSTEMS, INC. CONTACT INFORMATION
Orders for
Turner BioSystems' products may be placed by:
Phone: (408)
636-2400
Toll Free: (888) 636-2401 or
Fax: (408) 737-7919
Contact us
via our contact form
Internet: www.turnerbiosystems.com
Mailing Address: