A Modulus Microplate Luminometer Method for
Promega
ENLITEN® ATP Assay
INTRODUCTION
The Turner
BioSystems Modulus Microplate Luminometer combined with the Promega
ENLITEN® ATP Assay System Bioluminescence Detection Kit provides a
highly sensitive, quantitative method for rapidly measuring adenosine
5'-triphosphate (ATP). Some of the many applications for the Promega ENLITEN®
ATP Assay System include indirect measurement of bacteria, yeasts, fungi
and other microorganisms and assay of enzymes that produce or degrade
ATP.
The ATP-dependent oxidation of luciferin by luciferase produces light
measured by the Modulus Microplate Luminometer. When ATP is the
limiting factor in the luciferin oxidation reaction, the amount of light
produced is proportional to the ATP concentration of the sample.
The highly
sensitive Modulus Microplate Luminometer with injector can detect
attomole levels of ATP. The limit of detection for ATP with the Promega
ENLITEN® detection assay is 1 attomoles or 0.5 femtograms.

Figure 1. A standard curve is obtained using the
Modulus Microplate Luminometer and the Promega ENLITEN® ATP
Assay System. 100 µL rL/L Reagent was injected into sample wells
of a 96-well plate containing 10 µL of ATP Standard diluted in HEPES
buffer.
When
designing an ATP assay with the Promega ENLITEN® Kit, it is important
to remember several key aspects of the luciferase reaction.
First, rL/L
reacts optimally at 7.73 pH and 23 - 25°C. Salts and many nonionic
chemicals will impair light production. Therefore, exercise caution when
selecting buffers and ATP extractants for the sample preparations. Second,
check for ATP contamination by comparing the RLU values obtained from
the assay buffer and rL/L to those of ATP-Free Water.
There may
exist several different ATP stores within any given sample. For example,
in cell preparations, ATP may be present in the media. Treatment of cells
may alter the amount of ATP in the media.
To measure
the ATP content in microorganisms or cells, extract the ATP before analysis.
Trichloroacetic acid (TCA) is recommended to release ATP from cells and
inactivate ATP-degrading enzymes.
Determining the minimum amount of TCA is crucial because TCA inhibits
the luciferase reaction. A final concentration of 0.5 - 2.5% TCA is sufficient
for ATP extraction from bacteria and eukaryotic cells.
Standard
curves are useful tools for proper ATP analysis of samples. Prepare a
standard curve daily or whenever a new aliquot of the rL/L Reagent is
used. The standard curve should include varying concentrations of ATP
diluted in assay buffer. The composition of this buffer should be identical
to the buffer composition used in assay sample preparation. Minding these
considerations safeguard the highest possible accuracy of ATP analysis.
MATERIALS
REQUIRED
- Modulus
Microplate Multimode
Reader
- Modulus Microplate Luminescence Module
- Modulus Microplate Injector System (P/N 9300-061 for one injector,
P/N 9300-062 for two injectors)
- Promega ENLITEN® ATP Assay System Bioluminescence Detection Kit
for ATP
(P/N FF2000) containing one vial rLuciferase/Luciferin (rL/L) Reagent,
12 mL Reconstitution Buffer, one vial ATP Standard (1x10-7 M), 25 mL
ATP-Free Water
- 96-well, white microplates (E&K Scientific, EK-25075)
- P10, P200 pipette and pipette tips
- assay buffer
- 1.5
mL microfuge tubes
- nitrile, vinyl, or latex gloves
Note: Individuals sensitive to latex should use vinyl or nitrile gloves.
Storage
Conditions: The rL/L Reagent and Reconstitution Buffer must be stored
at -20°C prior to reconstitution. Store ATP Standard at -20°C.
EXPERIMENTAL PROTOCOL
1. Reagent
Preparation Recommendation
Note:
ATP contamination will cause erroneous results and increase background.
Wear gloves to prevent hand-transmitted ATP contamination during reagent
preparation and while performing the assay.
- Equilibrate
the sample buffer to room temperature.
- Lightly
tap the vial of rL/L Reagent before opening to ensure the lyophilized
material collects at the bottom of the vial.
- Transfer
the vial contents of rL/L Reconstitution Buffer into the vial of rL/L
Reagent.
- Replace
the stopper and slowly invert the vial several times to dissolve the
contents. Do not shake the reagent bottle.
- Allow
the reconstituted rL/L Reagent to stand at room temperature for one
hour before use.
Note:
Reconstituted rL/L Reagent may be kept for eight hours at room temperature.
Store at -20°C in single-use aliquots for long-term storage.
- Prepare
a 10-fold serial dilution of ATP Standard (1x10-7 M) in assay buffer.
Dilute ATP to 1x10-13 M.
2. Instrument
Set Up
- For ATP
decontamination, prime the injector with a 50% bleach (sodium hypochlorite)
solution. Allow it to sit for one hour before rinsing. Flush the injector
by ten times with sterile, distilled or deionized water. Proper luminometer
care is important for low-assay background and precise ATP measurements.
- Go to
Select Protocol from the Home screen and follow the protocol wizard
to select the preset Enliten ATP protocol. Enter the following: Luminescence;
at the Preset tab, select Enliten ATP; Finish.
Note:
If the instrument has been previously exposed to reagents containing
ATP, such as the Promega LAR Reagent, the injector / tubing may be contaminated.
It is preferable to use an injector/tubing set that has not been exposed
to LAR when injecting rL/L reagent. For example, if injector #1 has
been used for LAR injection, then use the #2 injector for ATP protocol.
- The Instrument
Control screen shows all the reading parameters: integration, each injector
and injection volume, delay time, and that all the wells are selected.
If desired, change the delay and integration time settings and save
the changed protocol under a different name in the User protocol folder.
The default parameters of the Enliten ATP protocol are 100 µL
rL/L Reagent injection; with two seconds delay and ten seconds integration
time.
- Select
wells to be read on the "Plate Map" according to how the samples
are loaded into the plate.
- Prepare
the injectors. Place the injector intake tubing into the bottle of rL/L
Reagent.
- Prime
the injector using the Setup icon on the Read screen of Instrument control.
Place the Waste Collection Tray in the instrument. Follow the step-by-step
injector set-up wizard to prime.
Note: If it is important to conserve the reagent, a clean container
(such as a cell/tissue culture plate) can be used for recovery.
- Open the
instrument door by using the Door icon on the touch screen. Place the
plate with A1 well at the top right corner of the microplate sample
tray. Close the door by using the Door icon.
- Select
the Start icon on the touch screen to start a read.
- Refer
to the on-screen Help topics, Quick Start Guide, or the Operating Manual
for detailed instructions.
3. ATP
Standard Curve
- Manually
add 10 µL of assay buffer or ATP of a different concentration
to a 96-well white microplate (E&K Scientific, EK-25075), according
to the selected map in previous steps.
- Run the
Enliten ATP protocol.
- Collect
the RLU data from measuring ATP sample concentrations from
1x10-13 - 1x10-7 M ATP.
4. Sample
Analysis
- Manually
add 10 µL of sample prepared in assay buffer to a 96-well white
microplate (E&K Scientific, EK-25075), according to the previously
selected map.
- Run the
Enliten ATP protocol.
- RLU values
measured by the Modulus Microplate Luminometer will appear on
the Results screen of the touch-screen display immediately after each
well is measured.
- Once the
measurements are complete, data can be transferred to an external computer
for further data analysis in Excel by using the provided USB flash drive.
- Remove
the plate after completion.
- Use the
Reverse Purge function in the Setup wizard to recover unused reagent,
if desired.
- Use the
Flush function in the Setup wizard to flush the injectors thoroughly
after use.
- Refer
to the Operating Manual for detailed instructions on injector use.
RESULTS
Sensitivity: 1x10-18 moles (1 attomole) of ATP using the Promega ENLITEN® ATP Assay
Dynamic Range: Up to six orders of magnitude dynamic range using
the Promega ENLITEN® ATP Assay
Minimum Detection Limit: 2x10-19 moles (0.2 attomole) of ATP using
Promega ENLITEN® ATP Assay
CONCLUSION
The Modulus
Microplate Luminometer offers superior sensitivity and dynamic range for
luminescence-based ATP assay. The Modulus Microplate luminometer achieves
its superior performance with a combination of unique detection and optical
designs, premium components such as the photomultiplier tube (PMT), low-noise
circuitry, and proprietary dual-masking system.
The modular
approach of the Modulus Microplate Luminometer allows for instrument
capability expansion as needs in the lab change. Fluorescence and/or absorbance
detection modules as well as other accessories can be added after the
initial purchase.
The superior
performance, ease of use, and utmost flexibility of the Modulus
Microplate make it an ideal microplate reader for today's life science
laboratory.
TURNER
BIOSYSTEMS CONTACT INFORMATION
Modulus is
a trademark of Turner BioSystems. Orders for Turner BioSystems' products
may be placed by:
Phone: (408)
636-2400
Toll Free: (888) 636-2401 or
Fax: (408) 737-7919
Contact us
via our contact form
Mailing Address:
Turner BioSystems, Inc.
645 N. Mary Avenue
Sunnyvale, CA 94085
TRADEMARKS
For research use only. Not for use in diagnostic procedures. Modulus is a trademark of Turner BioSystems, Inc. All other trademarks are the sole property of their respective owners.
ENLITEN is a trademark of Promega
Corporation and is registered with the U.S. Patent and Trademark Office. |