Veritas Microplate Luminometer Method for Promega Steady-Glo 1. INTRODUCTION The Turner BioSystems' Veritas Microplate Luminometer in combination with the Steady-Glo® Assay kit provides a convenient, rapid, and sensitive procedure for quantifying gene expression. Transcriptional regulation, coupled to the expression of a luciferase reporter gene, is regularly used to study a wide range of biological events in cultured cells. Luciferase is an ideal reporter because of the absence of endogenous luciferase activity in mammalian cells, and the functional enzyme is created immediately upon translation1,2. The Steady-Glo® Luciferase Assay System has been developed specifically to maximize the
sensitivity of the assay reagent while providing a luminescent signal
half-life of approximately five hours. The light signal can be measured
between 5 minutes and several hours after adding assay reagents. The extended
period of luminescence makes Steady-Glo® Luciferase Assay
System especially appealing to laboratory automation and high-throughput
applications. The Steady-Glo® Reagent is widely used in
the pharmaceutical and biotechnology industries. The Steady-Glo® Reagent is compatible with commonly used culture media for mammalian
cells (RPMI 1640, MEMa, DMEM and Ham's F12) and tolerates phenol red and
organic solvents. 2. MATERIALS REQUIRED
3. EXPERIMENT PROTOCOL 3.1 Reagent Preparation Steady-Glo® Substrate: Use as supplied. Store at -20°C, where it is stable for up to 6 months. The substrate may also be stored at 4°C for up to one month. Steady-Glo®Buffer: Use as supplied. Store below 25°C. Steady-Glo® Reagent: Transfer the contents of one bottle of Steady-Glo® Buffer to one bottle of Steady-Glo® Substrate. Mix by inversion until the substrate is thoroughly dissolved. Use reconstituted reagent on the same day it is prepared or store at -20°C for up to two weeks. Note: The temperature of the Steady-Glo® Reagent should be held constant at room temperature while quantifying luminescence, since luciferase activity is temperature dependent. Reagent stored frozen after reconstitution must be thawed below 25°C to ensure reagent performance. Mix well after thawing. The simplest method for thawing is placing the reagent in a water bath at room temperature. 3.2 Instrument Setup 3.2.1 Double click on the Veritas icon to start the software. 3.2.2 Click on "Run Promega Protocol" from the "Welcome to Veritas" dialog box. 3.2.3 Select "SteadyGlo" from the list of Promega protocols. 3.2.4 Enter your information into the "Experiment", "Operator", "Plate No.", and "Notes" fields in the "Main Dialog Box". 3.2.5 Click on "Options" from the "Main Dialog Box" and select the wells to be read, modify the number of runs, or set a delay time between each run. You can also modify the integration time in the "Other Options" tab. Once you have made your choices, click the "Apply Changes" button to return to the "Main Dialog Box". 3.3 Sample Analysis 3.3.1 Remove
the 96-well plate containing cell cultures from the incubator. 3.3.2 Add a volume of the Steady-Glo® Reagent equal to that of the culture medium in each well. For 96-well plates, typically 100 µL of reagent is added to cells grown in 100 µL of medium. 3.3.3 Wait a minimum of 5 minutes to allow for sufficient cell lysis, then proceed to measure with the Veritas Microplate Luminometer. 3.3.4 Insert
the plate into the Veritas Microplate Luminometer and
click "Start" to begin assay. RLU values measured by the Veritas Microplate Luminometer will appear in the Excel spreadsheet after all
the selected wells in each row have been read. If you encounter an error
message, refer to the troubleshooting guide for more information. 3.3.5 Once the measurements are complete, you can access Excel to analyze your data. 3.3.6 Please make sure to remove your plate after measurement. 4. REFERENCES 1. Ow, D.W. et al. (1986) Transient and stable expression of the firefly luciferase gene in plant cells and transgenic plants. Science 234, 8569. 2. De Wet, J.R. et al. (1987) Firefly luciferase gene: structure and expression in mammalian cells, Mol. Cell. Biol. 7, 72537. 5. ABOUT
PROMEGA CORPORATION
5. ABOUT
TURNER BIOSYSTEMS Phone: (408)
636-2400 Contact us
via our contact form Mailing Address: CAUTION: The lyophilized Steady-Glo® Substrate contains dithiothreitol (DTT) and is therefore classified as hazardous. The reconstituted reagent is not known to present any hazards as the concentration of DTT is less than 1%. However, we recommend the use of gloves, lab coats and eye protection when working with these or any chemical reagents. Promega and Turner BioSystems assume no liability for damage resulting from handling or contact with these products.
|
|