GloRunner™ Microplate Luminometer Method for

Alkaline Phosphatase

1. INTRODUCTION

The Turner BioSystems GloRunner™ Microplate Luminometer in combination with the APS-5 substrate from Lumigen, provides a quick, reliable method for measuring alkaline phosphatase activity. Alkaline phosphatase is a stable enzyme often conjugated to secondary antibodies for immunoassays. Alkaline phosphatase is also a useful genetic reporter.

The GloRunner detects as little as 3 x 10-20 moles alkaline phosphatase. Measurements are linear for over three orders of magnitude (Figure 1). All tests were conducted using Lumigen APS-5 (Lumigen, Southfield, MI) and purified calf intestinal mucosa alkaline phosphatase (Biozyme, San Diego, CA).


Figure 1: Detection of alkaline phosphatase with the GloRunner™ Microplate Luminometer.
90 µL APS-5 was added to 10 µL allkaline phosphatase diluted in dH2O. Samples were incubated at room temperature for 5 minutes before measurement.

2. MATERIALS REQUIRED

  • GloRunner™ Microplate Luminometer (P/N 9000-000)
  • 96-well plates, white (E&K Scientific EK 25075)
  • Lumigen APS-5
  • Alkaline phosphatase (Biozyme ALPI12G)
  • p200 pipette and pipette tips
  • p20 pipette and pipette tips

3. EXPERIMENT PROTOCOL

3.1 Reagent Preparation

  • Lumigen APS-5: Use as supplied. Store at 2–8°C, where it is stable for up to 12 months.
  • Alkaline phosphatase: Store at 4°C.

3.2 Instrument Setup

3.2.1 Double-click on the GloRunner icon to start the software.

    3.2.2 Open a new workbook and select the wells you wish to read from the "Options" screen.

    3.2.3 You may also wish to select a delay for incubation, the number of repeated runs, and a delay before microplate ejection. The recommended measurement duration is 1 second.

3.3 Sample Analysis

3.3.1 Equilibrate alkaline phosphatase, samples and APS-5 to room temperature.

3.3.2 Prepare a serial dilution of alkaline phosphatase in your assay buffer.

3.3.3 Add 10 µL of sample or standard to the 96-well plate.

3.3.4 Add 90 µL of APS-5 reagent to the sample or standard.

Note: The recommended ratio of sample volume to APS-5 volume is 1:10 and should not exceed 1:1

3.3.5 Incubate the microplate for 5 minutes at room temperature.

3.3.6 Insert the plate into the GloRunner sample tray and click on "Start" to begin assay.

3.3.7 The data will appear in the Excel spreadsheet after the GloRunner completes the run.

3.3.8 Please make sure to remove your plate after measurement.

4. ABOUT LUMIGEN

Orders for Lumigen's products may be placed by:
Phone: (248) 351-5600
Fax: (248) 351-0518
E-mail: marketing@lumigen.com
Internet: www.lumigen.com

Mailing address:
Lumigen, Inc.
22900 West Eight Mile
Southfield, MI 48034
USA

5. ABOUT BIOZYME

Orders for Biozyme's products may be place by:

Phone: (800) 423-8199
Fax: (858) 549-0138
E-mail: bioinfo@biozyme.com
Internet: www.biozyme.com

Mailing address:
Biozyme Laboratories International Ltd.
9939 Hibert Street, Suite 101
San Diego, CA 92131

6. ABOUT TURNER BIOSYSTEMS

Orders for Turner BioSystems' products may be placed by:

Phone: (408) 636-2400
Toll Free: (888) 636-2401 or
Fax: (408) 737-7919

Contact us via our contact form
Internet: www.turnerbiosystems.com

Mailing Address:
Turner BioSystems, Inc.
645 N. Mary Avenue
Sunnyvale, CA 94085

 

[Back to Top]


 
Promega Corporation