Modulus
Single Tube Fluorometer Method for
Quant-iT Protein
1. INTRODUCTION
The Modulus Fluorometer from Turner BioSystems in combination with
the Quant-iT Protein Assay kit from Molecular Probes provides an
accurate method for quantitation of protein in small volumes (100 µL).
The Quant-iT Protein assay is highly selective for protein and tolerates
common contaminants including salts, solvents or DNA, but not detergents.

Figure
1. BSA and Quant-iT Protein analyzed using the Modulus and the Blue
Fluorescence Optical Kit. 5 µL of each standard was added to 100
µL of Quant-iT RNA reagent. 100 µL of this mixture was transferred
to a minicell cuvette. The Modulus measured raw fluorescence of each dilution.
2. MATERIALS REQUIRED
- Modulus
Fluorometer (P/N 9200-000 or 9200-002)
- Blue Fluorescence
Optical Kit (P/N 9200-040)
- Quant-iT
Protein Assay Kit (Molecular Probes, Q33210)
- Minicell
Cuvettes (P/N 7000-950) and Minicell Adaptor (P/N 9200-928)
3. EXPERIMENTAL
PROTOCOL
3.1 Reagent
Preparation
NOTE: Handling, storage and use of the reagent should be performed
in accordance with the product information sheet supplied by Molecular
Probes, Inc.
The Quant-iT
Protein Reagent is supplied as a 1 mL concentrated dye solution in 1,2-propanediol.
On the day of the experiment, equilibrate kit contents to room temperature.
Prepare a working solution of the Quant-iT Protein Reagent by making a
1:200 dilution of the concentrated dye solution in Quant-iT Protein buffer.
Prepare this solution in a plastic container as the reagent may adsorb
to glass surfaces. Protect the working solution from light by covering
it with foil or placing it in the dark.
NOTE: For best results, use this solution within 3 hours of its
preparation.
3.2 Instrument
Set-Up
3.2.1 Power OFF the Modulus. Insert the Blue Fluorescence Optical
Kit and Minicell Adaptor according to Operating Manual.
3.2.2 Turn ON the Modulus. Allow a 5-minute warm up period.
3.3 Standard
Curve
3.3.1 Add 5 µL of each standard to a microcentrifuge tube containing
100 µL of the Quant-iT Protein Reagent working solution. Mix by
inversion.
3.3.2 Transfer 105 µL of each standard to a minicell cuvette.
3.3.3 The default mode for the Modulus is the Raw Fluorescence Mode.
Read all standards in the Raw Fluorescence Mode and record the readings.
3.3.4 Plot amount of the standard vs. fluorescence (FSU) and fit a
curve through the data points.
3.4 Sample
Analysis
3.4.1 Add 5 µL of each sample to a microcentrifuge tube containing
100 µL of the Quant-iT RNA reagent working solution. Mix by inversion.
3.4.2 Transfer 105 µL of each sample to a minicell cuvette.
3.4.3 Read each sample in the Raw Fluorescence Mode.
3.4.4 Use the standard curve plot from step 3.3.4 to determine
the amount of protein in the samples.
4. PATENTS
AND TRADEMARKS
Quant-iT is a registered trade--mark of Molecular Probes, Inc.
Modulus is
a trademark of Turner BioSystems, Inc.
5. ABOUT MOLECULAR PROBES, INC.
Orders for Molecular Probes' products may be placed by:
Phone: (541)
465-8338 or
Toll-Free: (800) 438-2209 (U.S. and Canada)
Fax: (541)
344-6504 or
Toll-Free: (800) 438-0228 (U.S. and Canada)
E-mail: order@probes.com
Mailing Address:
Molecular Probes, Inc.
PO Box 22010
Eugene, OR 97402-0414 USA
Information
on the scientific and technical back-ground of Molecular Probes' products
is available from the Technical Assistance Department:
Phone: (541)
465-8353
Fax: (541) 465-4593
E-mail: tech@probes.com
www.probes.com
6. ABOUT
TURNER BIOSYSTEMS, INC.
Orders for
Turner BioSystems' products may be placed by:
Phone: (408)
636-2400
Toll Free: (888) 636-2401 or
Fax: (408) 737-7919
Contact us
via our contact
form
www.turnerbiosystems.com
Mailing Address:
Turner BioSystems, Inc.
645 N. Mary Avenue
Sunnyvale, CA 94085
[Back
to Top] |