![]() |
Modulus Single Tube Fluorometer Method for DNA Quantitation using Hoechst Dye 33258 1. INTRODUCTION
3. FACTORS TO CONSIDER 3.1 The AT content of a DNA sample affects Hoechst 33258-DNA fluorescence. Hence, it is important to use a standard similar to the samples you are testing. Calf Thymus DNA can often serve as a reference for most plant and animal DNA because it is double-stranded, highly polymerized, and is approximately 58% AT (42% GC). For bacterial DNA, a different standard may be needed because the AT content varies widely depending on the species. 3.2 The conformation (supercoiled, relaxed, circular, linear) of plasmid DNA may result in different Hoechst 33258 binding efficiencies. Thus, it is important to select a standard with similar physical characteristics to your sample. 3.3 Hoechst 33258 fluoresces only about half as much when it binds to single-stranded genomic DNA compared to when it binds to double-stranded genomic DNA. In addition, short pieces of single-stranded DNA will not normally cause Hoechst 33258 to fluoresce in proportion to their concentration. 3.4 Buffers commonly used to extract DNA from whole cells have little or no effect on this assay. 3.5 Low levels of detergent (<0.01% SDS) have little or no effect on this assay. 3.6 Salt concentrations in the sample extract of up to 3 M NaCl do not affect this assay. For peak fluorescence, at least 200 mM NaCl is required for purified DNA, and 2.0 to 3.0 M is required for crude samples. In crude samples, higher salt concentrations appear to cause the dissociation of proteins from DNA, allowing the dye molecules to bind to DNA easier. 3.7 RNA does not interfere significantly with the DNA assay because Hoechst 33258 does not normally bind to RNA. Under high salt concentrations, fluorescence from RNA is usually less than 1% of the signal produced from the same concentration of DNA. 4. REAGENT PREPARATION NOTE: Hoechst 33258 is a possible carcinogen and possible mutagen. Wear gloves and a mask, and work under a fume hood. 4.1 Hoechst
33258 stock dye solution 4.2 10XTNE buffer stock solution: NOTE: The pH and NaCl concentration are essential for proper binding of the Hoechst reagent. 4.3 1XTNE: Dilute 10 mL 10XTNE with 90 mL distilled, 0.45 µm filtered water. 4.4 To prepare a 2X Dye Solution (200 ng/mL) for 10-1000 ng/mL final DNA concentration: 4.5 Calf thymus DNA standard: 5. Instrument
Set-Up 5.2 Protocol for 10 x 10 mm Standard Methacrylate Cuvettes 5.2.1 Prepare a 1-mL of 2000 ng/mL dsDNA standard solution for the calibration standard in a standard methacrylate cuvette. NOTE: Polystyrene cuvettes are not compatible with the UV Fluorescence Optical Kit. 5.2.2 Add 2XHoescht dye working solution at a 1:1 ratio to the 2000 ng/mL DNA standard. The final concentration is 1000 ng/mL. 5.2.3 In a separate cuvette, prepare a blank solution by adding 2XHoescht dye working solution at a 1:1 ratio with 1XTNE buffer. The minimum volume is 2 mL. 5.2.4 Calibrate the Modulus with 1000 ng/mL. NOTE: To optimize the accuracy, use a standard that is at or near the concentration of a typical sample. For example, if a typical sample is 300 ng/mL DNA, use a standard of 500 ng/mL DNA. The standard should be at or above 100 ng/mL. 5.2.5 Save the calibration for future use (optional) 5.2.6 Add 1 mL of 2XHoecsht dye to 1 mL of a sample in a methacrylate cuvette. If necessary, dilute the samples with 1XTNE. NOTE: The minimum volume is 2 mL in a 10 x 10 mm methacrylate cuvette. 5.2.7 Allow the samples and Hoechst dye a 5-minute equilibration period
before reading in the Modulus. NOTE: The final concentration is at least ½ of the original concentration due to the addition of the Hoechst dye. In addition, it is important to calculate for any dilutions of the original sample. 5.3 Protocol for Minicell Cuvettes 5.3.1 Prepare the standard solution. Dilute 1 mg/mL stock solution of DNA to a concentration of 2 µg/mL in 1XTNE. Add an equal volume of the 2XHoechst dye working solution, prepared in step 4.4. Mix well in a microcentrifuge tube. NOTE: To optimize the accuracy, use a standard that is at or near the concentration of a typical sample. For example, if a typical sample is 300 ng/mL DNA, use a standard of 500 ng/mL DNA.. The standard should be at or above 100 ng/mL. 5.3.2 Prepare the blank solution. Add an equal volume of the sample buffer
(usually 1XTNE) to 2XHoechst dye working solution in a separate microcentrifuge
tube. 5.3.4 Transfer 100 µL of the each sample, standard and blank solution to a minicell cuvette. Incubate for 2-5 minutes at room temperature, protected from light. NOTE: Do not introduce air bubbles in the minicell cuvette. Air bubbles
will cause erroneous readings. 5.3.6 Save the calibration for future use (optional). 5.3.7 Measure the sample solutions the concentration of DNA in the minicell cuvette will appear on the touchscreen. NOTE: The final concentration is at least ½ of the original concentration due to the addition of the Hoechst dye. In addition, it is important to calculate for any dilutions of the original sample. 6. ABOUT
TURNER BIOSYSTEMS Phone: (408)
636-2400 Contact us
via our contact
form Mailing Address: 7. ABOUT
MOLECULAR PROBES Phone: (541)
465-8338 or Fax: (541)
344-6504 or Mailing Address:
|
|
| Promega Corporation Sunnyvale Office • Detection Instruments 645 N Mary Avenue • Sunnyvale CA 94085 Phone: 408.636.2400 - Toll Free: 888.636.2401 - Fax: 408.737.7919 |
||
|
||